| Technical parameters |
Model NOVA8000 Series |
| Configuration |
Flame System |
Graphite System |
| Linearity Error (%) |
≤10 |
| Wavelength error |
±0.15nm |
| Wavelength repeatability |
≤ 0.05nm |
| Wavelength range |
(185~910) nm |
| Spectral bandwidth |
Spectral bandwidth 0.1, 0.2, 0.4, 1.0, 2.0nm five
automatically selectable.
|
| Photometer |
High luminous flux total reflection dual beam spectroscopy
system, single beam / dual beam any automatic switching.
|
| Baseline stability |
Zero-point drift: ± 0.002Abs/30 min; transient noise: ≤ 0.0025Abs |
| Resolution |
Mn 279.5 nm and 279.6 nm two spectral lines can be resolved, and the peak and valley energy values between the two lines is less than ≤ 25% of the energy |
| Turning lamp holder |
Eight-lamp position automatic conversion lamp holder, automatic collimation, 8-lamps can be preheated and lit at the same time. The lamp holder is designed to house the HCL and EDL lamps with Fully Automated Lamp selection, alignment and operation is software controlled. |
| Lamp Life |
>5000 hours |
| Lamp Recognition |
Lamps are assigned at the fixed positions and changed automatically through software selection. RFID is optional. |
|
Characteristic
concentration (Cu)
|
0.03μg/ml/1%. |
| Monochromator |
Czerny -Tuner Holographic grating & rotational (1,800 lines/mm) |
| Detection limit |
Flame method for copper measurement: ≤ 0.002 μg/mL |
Graphite Furnace method for Lead (Pb) measurement: ≤ 0.001 μg/mL(Cd < 0.3X10-12g |
| Repeatability |
RSD ≤ 0.25 %. |
| Combustion head |
Titanium metal combustion head. |
| Nebulizer |
High-efficiency glass, stainless steel nebulizer. |
| Atomizing chamber |
Explosion-proof and corrosion-resistant material atomizing chamber. |
| Graphite furnace |
Highly stable Flame Atomizer
Support Integrated Autosampler
|
Heated graphite furnace atomizer (HGA)
Integrated autosampler operation and furnace camera optional
|
| Graphite Furnace System Control |
Temperature ranges from ambient temperature to 3,000 °C, programmable in intervals of 0.2 °C to 3,000 °C/s
Real-time temperature control
Self-Test enabled system
|
| Detector |
High sensitivity Photo-Multiplier Tube (PMT) detector |
| Background Correction |
Automatic deuterium lamp and self-reversal deuterium correction: can correct 1A background, greater than or equal to 60 times the correction capacity. |
| Data Processing |
Measurement method: flame method, graphite furnace method Concentration calculation method: Standard curve method (1~2 times curve), auto-fit, standard addition method. Repeated measurement times: 1-99 times, calculate the average value, give standard deviation and relative standard deviation. Result printing: Parameter printing, data result printing, graphic printing. |